HALL EFFECT PROBES
MPT Model (Miniature)
LPT Model (Limpet)
QUICK FACTS
-
There are two main families of probes manufactured by Group3, the MPT and the LPT series probes
-
Every Group3 probe is individually characterized at many hundred field values
-
Any Group3 probe can be used with any Group3 Teslameter
-
Probe cables can be ordered in any length up to 30m
-
A printed calibration data of the individual characterization tables for that particular probe is supplied.
-
To be use in conjunction with a DTM-151 or DTM152 for temperature compensation and DTM-133 or DTM-333 for no temperature compensation.
Probe Ordering Guide
(4-Range: 0.3, 0.6, 1.2 , 3.0 Tesla )
​​
MPT - 1 4 1 - 2 S
Probe Sensor Head
Series Model
MPT - Miniature
LPT -Limpet
Cable Length
Shielding
Probe Length, meters
S - Shielded
M - No Shielding
141 - With Temperature Compensation, standard sensitivity ,
0.3, 0.6, 1.2, 3.0 Tesla Ranges
132 - No Temperature Compensation, standard sensitivity ,
0.3, 0.6, 1.2, 3.0 Tesla Ranges
231 - With Temperature Compensation, high sensitivity ,
0.03, 0.06, 0.12, 0.3 Tesla Ranges
230 - No Temperature Compensation, high sensitivity ,
0.03, 0.06, 0.12, 0.3 Tesla Ranges
130 - For LPT , No Temperature Compensation, standard sensitivity ,
0.3, 0.6, 1.2, 3.0 Tesla Ranges
Probe Ordering Guide
(Single Range Probes)
Choose from 0.3, 0.6, 1.2 , 3.0 Tesla for Standard Sensitivity
Choose from 0.03, 0.06, 0.12 , 0.3 Tesla for High Sensitivity
​​
MPT - 1 4 1 - 30 - 2 S
Probe Sensor Head
Series Model
MPT - Miniature
LPT -Limpet
141 - With Temperature Compensation, standard sensitivity ,
132 - No Temperature Compensation, standard sensitivity ,
231 - With Temperature Compensation, high sensitivity ,
230 - No Temperature Compensation, high sensitivity ,
130 - For LPT , No Temperature Compensation, standard sensitivity ,
Range
Cable Length
Shielding
Probe Length,
meters
S - Shielded
M - No Shielding
03 - 0.03 Tesla for High Sensitivity, 0.3 Tesla for Standard Sensitivity
06 - 0.06 Tesla for High Sensitivity, 0.6 Tesla for Standard Sensitivity
12 - 0.12 Tesla for High Sensitivity, 1.2 Tesla for Standard Sensitivity
30 - 0.3 Tesla for High Sensitivity, 3.0 Tesla for Standard Sensitivity
FEATURES
-
Probe cables can be ordered in any length up to 30 meters.
​
-
Unshielded cable has a diameter of 4.5 mm.
​
-
Shielded cable has a diameter of 6.3 mm
​
-
The probe head is usually on the end of a short (10 to 50 mm) length of twisted but unjacketed wires. This allows greater flexibility when mounting the probe head.
​
-
Various special configurations of probe head have been manufactured to special customer requirements - if one of the existing types of probe is not suitable for your application, please contact your Group3 representative.
​
-
The operating temperature range of the probe and instrument is 0 to 50 ° C
​
-
A range of probe holders is available for both series of probe heads. One, two and three axis probe holders have been manufactured in aluminium and non-metallic materials.
​
-
A printed booklet of the individual characterization tables for that particular probe is supplied.
SPECIFICATIONS
MPT Series
-
miniature head, epoxy resin-encapsulated
-
sensitive area 1.0 x 0.5 mm
-
housed in a probe head of 14 (L) x 5 (W) x 2 mm thick
For full accuracy over a range of temperatures the DTM-151 or DTM-152 should be used with the following probes:
MPT Series – 132, 132A, 141, 230, 231
MPT series with Standard Sensitivity
Probe Probe head size Active Area Max Field Accuracy Tempco Zero Drift
(mm) (mm) at 25° C (ppm /°C) (µT/°C)
​
MPT-141 14 x 5 x 2 1 x 0.5 3T ± 0.01% ± 5 ± 1
MPT-132 14 x 5 x 2 1 x 0.5 3T ± 0.03% - 80 ± 40
​
​
MPT series with High Sensitivity
Probe Probe head size Active Area Max Field Accuracy Tempco Zero Drift
(mm) (mm) at 25° C (ppm /°C) (µT/°C)
​
MPT-231 14 x 5 x 2 1 x 0.5 0.3T ± 0.03% ± 20 ± 1
MPT-230 14 x 5 x 2 1 x 0.5 0.3T ± 0.03% -600 ± 8
​
​
LPT Series
​
-
sensitive area 1.6 x 4.0 mm
-
housed in a probe head of 13 x 10.5 x 2.5 mm thick (LPT-130 and LPT-230)
-
14 x 14 x 2.5 mm thick (LPT-141 and LPT-231)
For full accuracy over a range of temperatures the DTM-151 or DTM-152 should be used with the LPT-141 and LPT-231.
LPT series performance standard sensitivity - 130, 141, 230, 231
LPT series with Standard Sensitivity
​
Probe Probe head size Active Area Max Field Accuracy Tempco Zero Drift
(mm) (mm) at 25° C (ppm /°C) (µT/°C)
​
LPT-141 14 x 14 x 2.5 4 x 1.6 3T ± 0.01% ± 5 ± 1
LPT-130 13 x 10.5 x 2.5 4 x 1.6 3T ± 0.03% -60 ± 12
LPT series with High Sensitivity
Probe Probe head size Active Area Max Field Accuracy Tempco Zero Drift
(mm) (mm) at 25° C (ppm /°C) (µT/°C)
LPT-231 14 x 14 x 2.5 4 x 1.6 0.3T ± 0.03% ± 20 ± 1
LPT-130 13 x 10.5 x 2.5 4 x 1.6 0.3T ± 0.03% -600 ± 8
​
CUSTOM-MADE PROBES
Group3 creates custom-made probes to suit your field measurement needs. Below are some of the customized probes we can create for you:
​
-
High field measurement of up to 8 Tesla.
-
Single range probe to any of the following:
-
Standard Sensitivity - 0.3 Tesla, 0.6 Tesla, 1.2 Tesla or 3.0 Tesla​
-
High Sensitivity - 0.03 Tesla, 0.06 Tesla, 0.12 Tesla or 0.3 Tesla
-
-
Probe isolation using circular MIL spec connectors.
-
Rotating probe head side using brush-less slip rings such as a MERCOTAC connector.
​