top of page

DTM-152 

QUICK FACTS

​​

  • Accuracy is 0.01% of reading + 0.006% of full scale when used with MPT-141 probe

  • 0.001G or 0.1uT Resolution at 0.3T range 

  • 4 ranges

  • Maximum field 3 Tesla when used with a standard sensitivity probe

  • 7 digit display on a 7"tablet

  • Data logging, no need for a separate computer

  • Control over desktop PC using LabView Interface provided 

  • Touch screen display

  • Communicate via LAN/Ethernet port

  • Temperature range for instrument and probe      0 to 50° C

  • To be used in conjunction with an MPT-141 or LPT-141 for standard sensitivity or an MPT-231or LPT-231 Hall Effect Probes for high sensitivity, low-field operation.

  • Mount as a bench top unit or in a standard 19" rack panel using mounting cheeks (ordered separately)

Ordering Guide

​​

D T M     -    1 5 2     -     D        N

Product Initials

DTM - Digital Teslameter

Series Model

Mount

Style

D - Bench Top

P - Panel Mount

Communication

152 - With Temperaure Compensation, 

         Touchscreen Display

​

N - Network

FEATURES

  • Because the probe characterisation information is stored in the probe assembly itself, any Group3 probe can be used with any Group3 DTM. However for full accuracy the DTM-152 should be used with a probe that has temperature compensation circuitry.​

​

  • Digital filtering (time averaging) can be enabled to suppress short term fluctuations.​

​

  • External DIP switches for device hardware address.

​

  • One (1) analog output are available. These are a buffered version of the raw probe signal, and are not corrected for linearity or temperature.

​

​​

​​

 SPECIFICATIONS

   DTM-152 performance summary when combined with the listed probe.

 

   Probe            Active area (mm)        Resolution of display        Max field         Finest resolution        Accuracy at 25°C        Tempco            Zero Drift   

                                                                                                                                                                                                            ppm /°C                 µT/°C

 

 

  LPT-141               4 x 1.61 in                       600,000                           3T                      1µT                                  ± 0.01%               ± 10                     ± 1

  LPT-231               4 x 1.61 in                       600,000                        0.3T                   0.1µT                                  ± 0.03%               ± 25                     ± 1

  MPT-141              1 x 0.51 in                       600,000                          3T                       1µT                                  ± 0.01%               ± 10                     ± 1

  MPT-231              1 x 0.51 in                       600,000                       0.3T                    0.1µT                                  ± 0.03%               ± 25                     ± 1

 

  The following specifications refer to DTM-151 with LPT-141 probe (standard size probe).

 

  DTM-152 WITH LPT-141 PROBE (LIMPET PROBE)

 

  Four ranges, selectable from front panel:

          1. 0.3
          2. 0.6
          3. 1.2
          4. 3.0    tesla full scale

 

   Note: Resolution depends on the range selected, and whether the digital filtering is enabled.

   Note: Resolution available at the serial port is greater than that of the display.

   

   

         DTM-152 with LPT-141 probe Filtering ON

 

                  Range selected                                                    Resolution (µT) Display                                                Resolution (µT) serial                                         

                     

                       0.3 tesla                                                                              0.1                                                                                       0.1

                       0.6 tesla                                                                              1                                                                                          1

                       1.2 tesla                                                                              1                                                                                          1

                       3.0 tesla                                                                              1                                                                                          1

 

 

   

         DTM-152 with LPT-141 probe Filtering OFF

 

                  Range selected                                                    Resolution (µT) Display                                                Resolution (µT) serial/GPIB                                           

                     

                     

                       0.3 tesla                                                                               5                                                                                       0.1

                       0.6 tesla                                                                              10                                                                                         1

                       1.2 tesla                                                                              20                                                                                         1

                       3.0 tesla                                                                              50                                                                                         1

​

 

 

 

         Absolute Accuracy

                  ± 0.01% of reading ± 0.006% of range at 25° C

 

        Temperature Stability

​

  •          scale factor: ± 5 ppm of reading / ° C typical, ± 10 ppm of reading / ° C max.

  •          zero drift: ± (1m T + 0.0003% of full scale ) / ° C max.

  •          Effect of probe cable: add -3ppm / ° C for each metre of probe cable.

 

 

         Frequency

                DC, or in AC mode 8 to 3,000 Hz

       

         Update rate

               3 fully corrected measurements per second

 

         Display

                7 digit, 7"tablet running Group3 App, interactive touchscreen display.  

 

         Enclosure

               All metal, with tilt stand on bench models, mounting cheeks on rack model

 

         Dimensions

               350 x 320 x 146 mm (including feet down)

     350 x 320 x 180 mm (including feet up)

​

Weight

     5.8 kg

Recommended Probes for DTM-152 Teslameters 

MPT-141

LPT-141

bottom of page