DTM-133 Series
Ordering Guide
​​
D T M - 1 3 3 - D S
Product Initials
DTM - Digital Teslameter
Series Model
Mount
Style
D - Bench Top
P - Panel Mount
151 - With Temperature Compensation
133 - No Temperature Compensation
152 - With Temperaure Compnesation,
Touchscreen Display
333 - No Temperature Compensation, 3-Axis,
3-Channel, Touch Screen Display
Communication
S - Serial RS232
G - GPIB, IEEE-488
​
FEATURES
-
Temperature coefficient is 80 ppm / ° C when used with LPT-130, or 140 ppm / ° C when used with MPT-132 probe.
​
-
Accuracy and temperature specifications include performance of probe.
​
-
Because the probe characterisation information is stored in the probe assembly itself, any Group3 probe can be used with any Group3 DTM.
​
-
Two buttons on the front panel are used to select the operation of the teslameter. The following modes can be selected for the display - magnetic field, peak hold of magnetic field, AC component of field.
​
-
Digital filtering (time averaging) can be enabled to suppress short term fluctuations.
​
-
Several teslameters with the serial communications option (up to 31) can be formed into a local communications loop, all of them talked to by the one serial port on the control computer. The system can be operated in a trigger mode, where the timing of the measurements by several teslameters can be synchronised
​
-
Internal switches select serial data format, baud rate, device address, string terminators, filtering, gauss or tesla units, data format, service request action, EOI action, and reset system to defaults.
​
-
An analog output is available. This is a buffered version of the raw probe signal, and is not corrected for linearity or temperature. It offers an instantaneous field value, (0 to 9kHz)
​
-
All models are available without display or front buttons, for true “Black Box” requirements.
​
-
Panel mount versions are available, and panels with correct cut-outs and mounting hardware.
​
-
The operating temperature range of the probe and instrument is 0 to 50 ° C
SPECIFICATIONS
DTM-133 performance summary when combined with the listed probe.
Probe Active area (mm) Resolution of display Max field Finest resolution Accuracy at 25°C Tempco Zero Drift
ppm /°C µT/°C
LPT - 130 4 x 1.61 1 in 120,000 3T 50µT ± 0.01% ± 80 ± 12
LPT - 230 4 x 1.61 1 in 12,000 0.3T 5µT ± 0.03% ± 620 ± 8
MPT - 132 1 x 0.51 1 in 120,000 3T 50µT ± 0.03% ± 140 ± 40
MPT - 230 1 x 0.51 1 in 12,000 0.3T 5µT ± 0.03% ± 620 ± 8
The following specifications refer to DTM-133 with LPT-130 probe (standard size probe).
DTM-133 WITH LPT-130 PROBE (LIMPET PROBE)
Four ranges, selectable from front panel:
1. 0.3
2. 0.6
3. 1.2
4. 3.0 tesla full scale
Note: Resolution depends on the range selected, and whether the digital filtering is enabled.
Note: Resolution available at the serial⁄GPIB port is greater than that of the display.
DTM-133 with LPT-130 probe Filtering ON
Range selected Resolution (µT) Display Resolution (µT) serial/GPIB
0.3 tesla 50 0.1
0.6 tesla 100 1
1.2 tesla 200 1
3.0 tesla 500 1
Absolute Accuracy
0.03% of reading ± 0.03% of range at 25° C
Temperature Stability with LPT-130 Probe
-
​ scale factor: -80 ppm of reading / ° C max.
-
Zero drift: ± (12m T + 0.0015% of full scale ) / ° C max.
-
Effect of probe cable: Add -3ppm / ° C for each metre of probe cable.
Temperature Stability withMPT-132 Probe
-
​ scale factor: ± (40m T + 0.0003% of full scale ) / ° C max.
-
zero drift: ± (15m T + 0.0010% of full scale ) / ° C typical. ± (40m T + 0.0015% of full scale ) / ° C max.
-
effect of probe cable: add -3ppm / ° C for each metre of probe cable
​
Frequency
DC, or in AC mode 8 to 3,000 Hz
Update rate
10 fully corrected measurements per second
Display
7 digit, red LED. 8 additional indicators for range, units, peak hold and filter
Enclosure
All metal, with tilt stand on bench models, bezel mount on rack models
Dimensions
217 x 125 x 50mm, weight 1.2 kg